Protein desalting column(5KMWCO,0.7ml)

Braveds
Description
Product Name Protein desalting column(5KMWCO,0.7ml)
Types Consumable material
Alias Protein centrifugal desalting column
Producer China
Application Desalting and replacement buffer
Storage 2~30℃
Synonyms Protein desalting column(1.5KMWCO,10ml) - B1K100
Protein desalting column(1.5KMWCO,8ml) - B1K080
Protein desalting column(1.5KMWCO,5ml) - B1K050
Protein desalting column(1.5KMWCO,0.7ml) - B1K007
Protein desalting column(5KMWCO,10ml) - B5K100
Protein desalting column(5KMWCO,8ml) - B5K080
Protein desalting column(5KMWCO,5ml) - B5K050
Overview

The protein desalination centrifugal column is an improved glucan desalination device, which can achieve rapid and excellent desalination and high recovery rate for proteins greater than 1.5K&5K MWCO. Trouble-free desalination and buffer exchange are provided for samples ranging from 100μL to 3mL.

High recovery rate - The low-binding resin greatly enhances the protein recovery rate

Fast - No need to screen or wait for proteins that rely on gravity to flow out

Economy - More economical and practical than other pre-assembled columns available on the market

Stable - reusable

Operation steps

Balance the desalination column

a. Remove the plug at the bottom of the centrifugal desalination column. First, slightly twist it in one direction, then twist it in the other direction, and then put it back into the collection tube. Centrifuge at 1000×g for 1 minute and pour out the protective solution in the collection tube.

b. Place the centrifuge column back into the collection tube, add 250μl of equilibrium solution (the buffer solution to be replaced after sample desalination) to the centrifuge column, let it stand until the equilibrium solution has fully permeated the packing material, then centrifuge at 1000×g for 1 minute. Repeat this process three times to remove the collection solution.

2. Sample processing

Place the centrifugal column into a clean EP tube, add 100μl to 200μl of the sample, let it stand for 3 minutes, then centrifuge at 1000×g for 1 minute. The solution in the EP tube is the desalinated sample.

3. Regeneration and preservation of the desalination column

Add 300μl of 0.5M NaOH solution to the used desalination column, let it stand for 2 minutes, centrifuge at 1000×g for 1 minute, and repeat twice. Then add 300μl of deionized water, let it stand for 2 minutes, centrifuge at 1000×g for 1 minute, and repeat this process three times. After processing the desalination column, add 20% ethanol solution, place a red plug at the lower end, tighten the cap, and store it at 2 to 30 degrees Celsius.

 

Precautions

1.The sample processing volume should be within the column processing range. Excessive processing will lead to incomplete desalination, while insufficient processing will result in a decrease in sample recovery rate.

2. The packing material may lose water and shrink in high-concentration alcohol solutions or saturated salt solutions. Do not pass the above solutions through the column.

3. When adding samples, try to add them evenly to the center of the tube.

4. When the sample concentration is too low, it is necessary to first concentrate the sample to the required volume or concentration.


Recommended maximum sample loading quantity:

Protein desalting column (0.7ml) 

200ul

Protein desalting column (5ml)

1.5ml

Protein desalting column (8ml)

2.4ml

Protein desalting column(10ml)

3ml

 